Cortisol and Cortisone in Extracted Female Human Urine on Raptor Biphenyl by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Cortisol and Cortisone in Extracted Female Human Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Cortisol-d40.75367.4121.0-
2.Endogenous cortisol0.76363.3120.991.1
3.Cortisone-d80.88369.4167.9-
4.Endogenous cortisone0.89361.3163.291.0
*Isobaric matrix peaks
image
LC_CF0697
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:40 °C
Sample
Diluent:Mobile phase A
Conc.: Calculated concentration is 105.1 ng/mL and 135.9 ng/mL for cortisol and cortisone, respectively
Inj. Vol.:10 µL
Mobile Phase
A:Water, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.67030
1.000.67030
1.010.60100
1.500.60100
1.510.67030
3.000.67030
Max Pressure:400 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Method:
1. Centrifuge female human urine for 5 min at 4,500 rpm, 10 °C.
2. Aliquot 380 µL supernatant. Add 20 µL each of internal standard solution (1 µg/mL in methanol).
3. Load 200 μL of sample on to ISOLUTE SLE+ 200 µL supported liquid extraction plate (part# 820-0200-P01).
4. Apply a pulse of vacuum to initiate flow.
5. Wait 5 min for sample to completely absorb.
6. Extract samples with 1 mL of MTBE. Allow solvent to flow for 5 min under gravity. Apply vacuum for 10-30 sec to complete elution.
7. Evaporate extracts to dryness under a stream of nitrogen.
8. Reconstitute in 200 µL mobile phase A prior to analysis.
9. Vortex to mix.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us