Steroid Panel Analysis on Force Biphenyl
This chromatogram may contain analytes that are synonyms of your search terms.

Steroid Panel Analysis on Force Biphenyl

PeakstR (min)Q1Q3 QuantifierQ3 Qualifier
1.Cortisol2.3363.3121.2327.2
2.Estradiol2.6255.3159.1133.1
3.11-Deoxycortisol4.0347.4109.297.1
4.DHEA4.3271.3253.2213.2
5.Boldenone4.4287.2121.2135.2
6.Corticosterone4.4347.4329.2121.1
7.Stanozolol4.4329.381.195.1
8.Epitestosterone4.5289.2109.297.1
9.17-OH-progesterone4.8331.397.2109.5
10.Testosterone5.0289.2109.197.0
11.Trenbolone5.1303.3109.297.1
12.Methyltestosterone5.1271.2253.1199.2
13.Nandrolone5.3275.2109.2257.3
14.25-Hydroxyvitamin D36.1383.3365.3107.1
15.Androstenedione6.2287.297.1109.2
16.Progesterone7.5315.3109.297.1
image
LC_CF0702
ColumnForce Biphenyl LC Column (cat.# 9629512)
Dimensions:100 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:30 °C
SampleSteroids panel
Diluent:50:50 Methanol:water
Conc.:5 ng/mL
Inj. Vol.:50 µL
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.43070
8.000.40100
8.010.43070
10.000.43070
Max Pressure:250 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:Scheduled MRM
InstrumentUHPLC
Notes
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

steroid

,

biphenyl

,

ultra biphenyl

,

ms/ms

,

lc-ms/ms

,

hplc

,

triple quad

,

quadrupole

,

phenyl

,

ESI

,

ESI+

,

turboionspray

,

4000

,

API 4000

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us