Statins on Raptor ARC-18 (1.8 μm) by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Statins on Raptor ARC-18 (1.8 μm) by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonESI Mode
1.Pitavastatin0.6720.1422.20290.25+
2.Pravastatin0.8055423.25321.20-
3.Rosuvastatin1.1781482.25258.30+
4.Cerivastatin1.3850.1460.30356.30+
5.Fluvastatin1.6395412.25224.25+
6.Atorvastatin1.7131559.30440.20+
7.Mevastatin1.9641413.15311.35+
8.Lovastatin2.1291427.15325.25+
9.Simvastatin2.3161441.25325.40+
Pravastatin is in negative mode
image
LC_CF0710
ColumnRaptor ARC-18 (cat.# 9314232)
Dimensions:30 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:40 °C
Sample
Diluent:75:25 Water:acetonitrile
Conc.:0.1-5 ng/mL
Inj. Vol.:1 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.77525
2.500.72575
2.800.72575
2.810.77525
3.500.77525
DetectorMS/MS
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

statins

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us