Methylmalonic Acid in SeraFlx BIOMATRIX on Force C18 by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Methylmalonic Acid in SeraFlx BIOMATRIX on Force C18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion
1.Succinic acid1.69117.373.1
2.Methyl-D3-malonic acid2.02120.276.2
3.Methylmalonic acid2.03117.373.1
image
LC_CF0717
ColumnForce C18 (cat.# 963431E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:3 µm
Pore Size:100 Å
Guard Column:Force C18 EXP guard column cartridge 3 mm ID, (cat.# 963450253)
Temp.:35 °C
Sample
Conc.: 50 ng/mL methylmalonic acid in SeraFlx BIOMATRIX
Inj. Vol.:3 µL
Mobile Phase
A:0.5% Formic acid in water
B:0.5% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.7955
0.500.7955
3.000.7595
3.010.7955
5.000.7955
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC
Notes100 μL of sample (SeraFlx BIOMATRIX) was aliquoted for extraction. The sample was fortified at 50 ng/mL with methylmalonic acid. It was also fortified with 5 μL of succinic acid (20 μg/mL in water, for validation purposes only). 5 μL of the internal standard (2,500 ng/mL MMA-D3 in water) was added to the sample. The sample was precipitated using 300 μL of 0.5% formic acid in methanol followed by a 10 second vortex at 3,000 rpm. The sample was then centrifuged at 4,000 rpm for 10 minutes at 10 °C. 250 μL of the supernatant was filtered using a Thomson SINGLE StEP standard filter vial (PVDF, 0.2 μm, Restek cat.# 25895) prior to analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

MMA

,

B12

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us