Vitamin K1 and K2 in Fortified Artificial Serum on Raptor Biphenyl by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Vitamin K1 and K2 in Fortified Artificial Serum on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Vitamin MK41.440.5445.5187.2
2.Vitamin K11.470.5451.5187.2
3.Vitamin MK72.200.5649.7187.2
image
LC_CF0727
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:15:85 Water:methanol
Conc.:0.5 ng/mL Fortified SeraFlx BIOMATRIX sample
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.1% formic acid, 5 mM ammonium formate
B:Methanol, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.41090
1.000.40100
3.000.40100
3.010.41090
4.000.41090
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure:
A 500 µL aliquot of fortified sample (SeraFlx BIOMATRIX) was mixed with 5 µL of internal standard solution (K1-d7, MK4-d7, and MK7-d7 at 100 ng/mL in methanol) and 1.5 mL of acetonitrile followed by vortexing for 20 seconds at 3000 rpm. After centrifugation at 4300 rpm for 10 minutes, the supernatant was loaded onto a Biotage ISOLUTE PLD+ 96-well plate (50 mg) and vacuum was applied to collect the eluate. The eluate was then evaporated to dryness at 50 °C under a gentle stream of nitrogen. The dried extract was reconstituted with 100 μL of diluent and 5 μL of sample was injected for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

vitamin K1

,

vitamin

,

MK4

,

MK7

,

phylloquinone

,

menaquinone

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us