Phosphatidylethanol 16:0/18:1 in Human Whole Blood on Raptor FluoroPhenyl
This chromatogram may contain analytes that are synonyms of your search terms.

Phosphatidylethanol 16:0/18:1 in Human Whole Blood on Raptor FluoroPhenyl

PeakstR (min)Conc.Precursor IonProduct Ion
1.Phosphatidylethanol-d5 (PEth-d5) 16:0/18:11.580.4 μM706.6281.3
2.Phosphatidylethanol (PEth) 16:0/18:11.580.5 μM701.5255.3
image
LC_CF0733
ColumnRaptor FluoroPhenyl (cat.# 9319A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Conc.:0.5 μM in blood
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:9:1 Methanol:2-propanol
Time (min)Flow (mL/min)%A%B
0.000.53070
1.000.52080
2.000.52080
2.500.50100
2.510.53070
3.500.53070
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure (Provided by Redhot Diagnostics):
The fortified blood sample (50 µL) was mixed with 50 µL of internal standard (0.4 µM PEth-d5 in 2-propanol) and 150 µL of 4:1 2-propanol:tetrahydrofuran. The mixture was vortexed for 20 seconds at 3000 rpm and centrifuged for 10 minutes at 4300 rpm. The supernatant (2 µL) was injected for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PEth

,

phosphatidylethanol

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us