Carbamate Pesticides on Ultra Carbamate by LC-MS/MS

Carbamate Pesticides on Ultra Carbamate by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Aldicarb sulfone1.77240.186.2148.2
2.Aldicarb sulfoxide1.92207.1132.189.1
3.Oxamyl2.01237.172.190.1
4.Methomyl2.22163.1106.088.1
5.3-Hydroxycarbofuran2.68238.1163.0181.0
6.Aldicarb3.44208.1116.089.0
7.Propoxur3.71210.1111.0168.1
8.Carbofuran3.80222.1165.1123.0
9.Carbaryl4.07202.1145.0127.0
10.Methiocarb4.78226.1169.1121.1
image
LC_EV0534
ColumnUltra Carbamate (cat.# 9177312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:30 °C
Sample531.1 Carbamate pesticide calibration mixture (cat.# 32273)
Diluent:Water
Conc.:10 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid + 5 mM ammonium formate in water
B:0.1% Formic acid + 5 mM ammonium formate in methanol
Time (min)Flow (mL/min)%A%B
0.000.57525
5.000.5595
5.010.57525
8.000.57525
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us