share:
delim
delim
EPA 8310 PAHs on Pinnacle II PAH by LC-UV
This chromatogram may contain analytes that are synonyms of your search terms.

EPA 8310 PAHs on Pinnacle II PAH by LC-UV

PeakstR (min)Conc.
(µg/mL)
1.Naphthalene2.68100
2.Acenaphthylene3.09100
3.1-Methylnaphthalene3.39100
4.2-Methylnaphthalene3.56100
5.Acenaphthene3.74100
6.Fluorene3.91100
7.Phenanthrene4.40100
8.Anthracene4.93100
9.Fluoranthene5.4910
PeakstR (min)Conc.
(µg/mL)
10.Pyrene5.9110
11.Benzo[a]anthracene7.3910
12.Chrysene7.7710
13.Benzo[b]fluoranthene8.9510
14.Benzo[k]fluoranthene9.525
15.Benzo[a]pyrene9.9910
16.Dibenzo[a,h]anthracene10.7510
17.Benzo[ghi]perylene11.1810
18.Indeno[1,2,3-cd]pyrene11.6010
image
LC_EV0537
ColumnPinnacle II PAH (cat.# 921946E)
Dimensions:150 mm x 3.0 mm ID
Particle Size:4 µm
Pore Size:110 Å
Temp.:30 °C
SampleEPA Method 8310 quality control check (cat.# 31843)
Inj. Vol.:10 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.25050
10.01.20100
12.51.20100
12.511.25050
15.01.25050
DetectorUV/Vis @ 280 nm
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAH

,

8310

,

610

,

Polycyclic

,

aromatic

,

hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us