share:
delim
delim
Perfluorinated Alkyl Acids on Raptor C18 by LC-MS/MS for EPA Method 537

Perfluorinated Alkyl Acids on Raptor C18 by LC-MS/MS for EPA Method 537

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanesulfonic acid (PFBS)1.9810299.1079.99
2.Perfluoro-n-[1,2-13C2]hexanoic acid (13C-PFHxA)2.9210315.13270.13
3.Perfluorohexanoic acid (PFHxA)2.9310313.10269.12
4.Perfluoroheptanoic acid (PFHpA)4.005363.16319.09
5.Perfluorohexanesulfonic acid (PFHxS)4.1410399.1379.98
6.Perfluoro-[1,2-13C2]octanoic acid (13C-PFOA)4.835415.13370.10
7.Perfluorooctanoic acid (PFOA)4.835413.16369.10
8.Perfluorononanoic acid (PFNA)5.505463.16419.19
9.Perfluoro-1-[1,2,3,4-13C4]octanesulfonic acid (13C-PFOS)5.545503.1379.98
10.Perfluorooctanesulfonic acid (PFOS)5.5410499.1779.98
11.Perfluoro-n-[1,2-13C2]decanoic acid (13C-PFDA)6.065515.17470.17
12.Perfluorodecanoic acid (PFDA)6.075513.17469.16
13.N-deuteriomethylperfluoro-1-octanesulfonamidoacetic acid (d3-NMeFOSAA)6.285573.23419.15
14.N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA)6.285570.20419.17
15.N-deuterioethylperfluoro-1-octanesulfonamidoacetic acid (d5-NEtFOSAA)6.525589.27419.11
16.N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA)6.525584.20419.18
17.Perfluoroundecanoic acid (PFUnA)6.565563.23519.24
18.Perfluorododecanoic acid (PFDoA)6.985613.23569.19
19.Perfluorotridecanoic acid (PFTrDA)7.355663.23619.21
20.Perfluorotetradecanoic acid (PFTA)7.675713.23669.23
image
LC_EV0548
ColumnRaptor C18 (cat.# 9304552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:96:4 Methanol:water
Conc.:5–10 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:5 mM ammonium acetate in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47525
8.000.41585
8.010.47525
9.000.47525
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI-
Mode:MRM
InstrumentWaters ACQUITY
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Raptor C18

,

EPA 537

,

perfluorinated chemical

,

fluorochemical

,

Raptor

,

PFOA

,

PFOS

,

perfluorinated

,

perfluorinated alkyl acid

,

PFAS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us