share:
delim
delim
Ultra-short Chain, Legacy, and Alternative PFAS on Raptor C18 in Reagent Water

Ultra-short Chain, Legacy, and Alternative PFAS on Raptor C18 in Reagent Water

PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion
1.Perfluoropropanoic acid (PFPrA)2.7480162.9119.0
2.Perfluorobutanoic acid (PFBA)4.6980212.8169.0
3.Perfluoropropanesulfonic acid (PFPrS)5.1380248.879.6
4.Perfluorobutanesulfonic acid (PFBS)6.1480298.879.9
5.Perfluoro-n-[1,2-13C2]hexanoic acid (13C2-PFHxA)6.7550314.9270.0
6.Hexafluoropropylene oxide-dimer acid (HFPO-DA)6.9280285.0168.9
7.Ammonium 4,8-dioxa-3H-perfluorononanoate (ADONA)7.3380376.9250.7
8.Perfluorooctanoic acid (PFOA)7.7080413.1368.9
9.Perfluoro-[1,2-13C2]octanoic acid (13C2-PFOA)7.7050415.0370.0
10.Perfluorooctanesulfonic acid (PFOS)8.0180498.880.0
11.Perfluoro-[1,2,3,4-13C4]octanesulfonic acid (13C4-PFOS)8.0150503.080.0
12.9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate (9Cl-PF3ONS)8.1580530.8350.7
13.11-Chloroeicosafluoro-3-oxanonane-1-sulfonate (11Cl-PF3OUdS)8.6180630.7451.0
image
LC_EV0555
ColumnRaptor C18 (cat.# 9304A1E)
Dimensions:100 mm x 3 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Conc.:80 ppt
Inj. Vol.:10 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.258020
7.000.25595
9.000.25595
9.010.258020
11.00.258020
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
NotesA PFAS delay column (cat.# 27854) was installed between the pump mixer and the injector.

In a polypropylene vial, 250 µL of reagent water (fortified at 80 ppt) was mixed with 250 µL of 40:60 reagent water:methanol and 5 µL of internal standard solution (5 ng/mL of 13C2-PFHxA, 13C2-PFOA, 13C4-PFOS in methanol). The vial was capped with a polyethylene cap prior to analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Ultrashort Chain

,

PFPrA

,

PFPrS

,

ADONA

,

9Cl-PF3ONS

,

11Cl-PF3OUdS

,

PFAS

,

HFPO-DA

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us