Carbamates in Orange Oil on Ultra Carbamate

Carbamates in Orange Oil on Ultra Carbamate

Peaks
1.Aldicarb sulfone
2.Aldicarb sulfoxide
3.Oxamyl
4.Methomyl
5.3-Hydroxycarbofuran
6.Aldicarb
7.Propoxur
8.Carbofuran
9.Carbaryl
10.Methiocarb
11.BDMC (IS)
image
LC_FF0472
ColumnUltra Carbamate (cat.# 9177352)
Dimensions:50 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:ambient
Sample531.1 Carbamate Pesticide Calibration Mixture (cat.# 32273)
4-bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC) (cat.# 32274)
Diluent:methanol
Conc.:10 ppm spiked into unprocessed orange oil
Inj. Vol.:3 µL
Mobile Phase
A:2 mM ammonium acetate:methanol, 90:10
B:2 mM ammonium acetate:methanol, 10:90
Time (min)%B
0.0020
20100
25100
Flow:0.2 mL/min
DetectorLECO Unique® TOFMS
Ionization Source Type:high flow ESI
Ion Mode:positive
Nebulizing Pressure:100 kPa
Desolvation Gas (N2):4 L/min
Interface Temp.:120 °C
Nozzle Voltage:62 V
Capillary Voltage:2.75 kV
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

orange

,

orange oil

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us