share:
delim
delim
Pesticides on Pinnacle DB Aqueous C18
This chromatogram may contain analytes that are synonyms of your search terms.

Pesticides on Pinnacle DB Aqueous C18

PeakstR (min)PolarityMol. Wt. Transition 1
1.Methamidophos0.45positive141.13142.0→94.0
2.Acephate0.62positive183.17184.0→142.8
3.Bentazone0.77negative240.28239.0→131.8
4.Omethoate1.07positive213.14214.0→124.9
5.Bromoxynil1.32negative276.91275.9→80.8
6.Butoxycarboxin1.64positive222.27223.1→86.1
7.Aldicarb sulfoxide2.03positive206.26207.1→132.1
8.Methomyl2.08positive162.21163.1→88.1
9.Ioxynil2.46negative370.92369.8→126.7
10.Thiamethoxam2.7positive291.7292.0→211.0
11.Monocrotophos2.73positive223.17224.1→127.0
12.Dimethoate2.9positive229.26230.1→199.0
13.Fenuron3.1positive164.21165.1→72.1
14.Dioxacarb3.12positive223.23224.0→123.0
15.Dicrotophos3.28positive237.19238.0→112.0
16.Vamidothion3.45positive287.34288.0→146.0
17.Carbendazim3.73positive191.19192.0→160.0
18.Aldicarb3.73positive190.27208.1→116.0
19.Formetanate3.74positive221.26222.1→165.2
20.Acetamiprid3.95positive222.67223.1→126.1
21.Carbetamide3.99positive236.27237.0→192.0
22.Oxamyl3.99positive219.26237.1→72.2
23.Aminocarb4.11, 4.44positive208.26209.2→137.1
24.Propoxur4.15positive209.24210.1→111.0
25.Oxadixyl4.21positive278.31279.0→219.0
26.Bendiocarb4.24positive223.23224.1→109.1
27.Thiacloprid4.25positive252.71253.2→126.1
28.Thiophanate methyl4.3positive342.4343.0→151.0
29.Methabenzthiazuron4.31positive221.28222.0→165.1
30.Tricyclazole4.35positive189.23190.0→163.0
31.Pyracarbolid4.38positive217.27218.2→125.0
32.Propamocarb4.41positive188.27189.2→102.2
33.Carboxin4.51positive235.3236.0→143.0
34.Ethiofencarb4.52positive225.31226.1→106.9
35.Carbaryl4.58positive201.22202.1→145.1
36.Thiabendazole4.6positive201.25202.0→175.0
37.Carbanilide4.6positive212.25213.1→94.0
38.Clethodim4.61positive359.91360.0→164.0
39.Tebuthiuron4.68positive228.32229.2→172.4
40.Fluometuron4.75positive232.21233.1→72.0
41.Metobromuron4.82positive259.1259.0→170.2
42.Flutriafol4.84positive301.3302.0→123.0
43.Chlortoluron4.92positive212.68213.1→72.2
44.Simetryn4.96positive213.31214.0→124.0
45.Pirimicarb4.97positive238.29239.2→72.1
46.Forchlorfenuron5.04positive247.68248.0→129.1
47.Siduron5.09positive232.32233.3→137.2
48.Ethofumesate5.11positive286.35304.0→121.0
49.Cycluron5.13positive198.31199.1→89.1
50.Diuron5.13positive233.1233.1→72.0
51.Diethofencarb5.15positive267.33268.0→226.0
52.Furalaxyl5.18positive301.34302.2→95.0
53.Methiocarb5.24positive225.31226.1→169.2
54.Secbumeton5.24positive225.29226.2→169.9
55.Terbumeton5.25positive225.29226.2→169.9
56.Prometon5.26positive225.29226.2→142.3
57.Paclobutrazol5.26positive293.8294.0→70.0
58.Promecarb5.27positive207.27208.2→109.2
59.Dimoxystrobin5.27positive326.39327.1→205.0
60.Ametryn5.29positive227.33228.2→186.2
61.Methoprotryne5.32positive271.39272.2→240.2
62.Triadimefon5.32positive293.75294.0→197.0
63.Mexacarbate5.34positive222.28223.2→166.2
64.Fludioxonil5.34negative248.19247.0→179.9
65.Boscalid5.36negative343.21343.0→307.0
66.Azoxystrobin5.36positive403.3404.1→372.1
67.Methoxyfenozide5.38positive368.47369.0→149.0
68.Cyproconazole5.39positive291.78292.2→70.2
69.Prothioconazole5.41negative344.26341.9→305.8
70.Flufenacet5.46positive363.34364.0→152.0
71.Butafenacil5.47positive474.82492.2→331.1
72.Myclobutanil5.51positive288.78289.0→70.0
73.Dimethomorph5.52,5.54positive387.86388.0→301.0
74.Bromuconazole 475.54positive377.06378.0→159.0
75.Prometryn5.55positive241.36242.2→158.1
76.Fipronil5.56negative437.15434.9→329.8
77.Triticonazole5.58positive317.81318.0→70.0
78.Tetraconazole5.6positive372.12372.0→159.0
79.Uniconazole5.62positive291.78292.2→70.2
80.Fluoxastrobin5.62positive458.83459.2→427.2
81.Cyazofamid5.63positive324.78325.0→108.0
82.Terbutryn5.65positive241.36242.2→186.0
83.Tebufenozide5.65positive352.47353.1→133.1
84.Chloroxuron5.67positive290.75291.0→72.0
85.Picoxystrobin5.67positive367.25368.0→145.0
86.Carfentrazone ethyl5.69positive412.17412.0→346.0
87.Diclobutrazol5.71positive328.21328.0→70.0
88.Fenoxycarb5.72positive301.34302.0→88.1
89.Diflubenzuron5.72positive310.67311.2→158.1
90.Neburon5.74positive275.18275.0→88.0
91.Flusilazole5.76positive315.4316.0→247.0
92.Epoxiconazole5.76positive329.76330.0→121.0
93.Fenbuconazole5.76positive336.82337.0→125.0
94.Bupirimate5.77positive316.41317.0→166.0
95.Zoxamide5.77positive336.54336.2→187.1
96.Rotenone5.79positive394.42395.0→213.0
97.Benalaxyl5.83positive325.41326.0→148.0
98.Tebuconazole5.84positive307.82308.0→70.0
99.Penconazole5.93positive284.19284.0→159.0
100.Cyprodinil5.96positive225.29226.0→93.0
101.Propiconazole5.96positive342.22342.0→159.0
102.Alanycarb5.98positive399.52400.1→238.2
103.Triflumuron5.99positive358.68359.1→156.1
104.Thiobencarb6positive257.78258.1→125.0
105.Pyraclostrobin6positive387.83388.0→194.0
106.Benzoximate6.01positive363.8364.0→199.0
107.Pinoxaden6.01positive400.52401.3→317.2
108.Clofentezine6.11positive303.15303.0→138.0
109.Trifloxystrobin6.12positive408.38409.2→186.0
110.Prochloraz6.21positive376.67376.0→308.0
111.Difenoconazole6.29positive406.28406.0→251.0
112.Benfuracarb6.29positive410.53411.2→195.1
113.Triflumizole6.38positive345.75346.1→278.1
114.Buprofezin6.4positive305.44306.0→201.0
115.Piperonyl butoxide6.42positive338.45356.2→177.2
116.Pyriproxyfen6.52positive321.28322.0→96.0
117.Propargite6.68positive350.48368.0→231.0
118.Etoxazole6.87positive359.42360.1→141.0
119.Amitraz6.89positive293.41294.2→163.3
120.Noviflumuron7.08negative529.1527.1→343.8
121.Pyridaben7.18positive364.94365.0→147.0
122.Fenpyroximate7.2positive421.5422.0→366.0
123.Fenpropimorph7.38positive303.49304.0→147.0
124.Spiroxamine7.63,7.81positive297.48298.0→144.0
125.Spinosyn A8.37positive731.97732.6→142.2
image
LC_FF0479A
ColumnPinnacle DB Aqueous C18 (cat.# 9418252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.9 µm
Pore Size:140 Å
Temp.:35 °C
Samplemulticomponent pesticide standard
Diluent:water
Conc.:33.3 ppb each pesticide
Inj. Vol.:5 µL
Mobile Phase
A:10 mM NH4OAc in water
B:10 mM NH4OAc in methanol
Time (min)%B
0.0010
110
890
1090
1110
Flow:0.60 mL/min
Max Pressure:~517 bar
DetectorApplied Biosystems/MDS Sciex LC/MS/MS
Model #:4000 QTRAP LC-MS/MS system
Ion Source:TurboIonSpray®
Ion Spray Voltage:5 kV
Gas 1:40 psi (275.8 kPa)
Gas 2:60 psi (413.7 kPa)
Source Temp.:500 °C
InstrumentShimadzu UFLCXR
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

pesticide

,

multiresidue

,

UHPLC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us