Incurred Pesticides in Red Bell Pepper QuEChERS Extract on Ultra Aqueous C18
This chromatogram may contain analytes that are synonyms of your search terms.

Incurred Pesticides in Red Bell Pepper QuEChERS Extract on Ultra Aqueous C18

Peaks
1.Methamidophos
2.Omethoate
3.Thiamethoxam
4.Dimethoate
5.Clothianidin
6.Imidacloprid
7.Propamocarb
8.Carbaryl
9.Metalaxyl
10.Diphenylamine
11.Myclobutanil
12.Spinosyn A
13.Spinosyn D
image
LC_FF0515
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
SampleQuEChERS extract of red bell pepper
Nicarbazin (bis-nitrophenol urea) (cat.# 33261)
Diluent:Deionized water:acetonitrile (90:10)
Inj. Vol.:10 µL
Mobile Phase10 mM ammonium acetate in water:10 mM ammonium formate in methanol with gradient program
Flow:0.5 mL/min
DetectorAB SCIEX API 4000 LC/MS/MS System
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Mode:MRM
InstrumentAPI LC-MS/MS
NotesInternal Standard: 100 µL of QuEChERS internal standard mix for GC-MS analysis (cat.# 33267) and 100 µL of QuEChERS internal standard mix for LC-MS/MS analysis (cat.# 33261) were added to each 10 g sample. These internal standard mixes require no dilutions (“snap-and-shoot”) and contain compounds specified in the EN 15662 QuEChERS method.

Extraction: 10 mL acetonitrile was added to a 10 gram sample of homogenized red bell pepper. After a 1 minute shake, Q-sep European method extraction salts (cat.# 26236) were added followed by another 1 minute shake. The sample was then centrifuged with a Q-sep 3000 centrifuge (cat.# 26230).

dSPE: Add 1 mL extract to QuEChERS dSPE cleanup tube containing 150 mg MgSO4, 25 mg PSA, and 25 mg C18 (cat.# 26216). Shake for 30 seconds and centrifuge with Q-sep 3000 centrifuge (cat.# 26230).
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us