LC Multiresidue Pesticide Standard #4 on Ultra Aqueous C18 by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

LC Multiresidue Pesticide Standard #4 on Ultra Aqueous C18 by LC-MS/MS

PeakstR (min)Precursor Ion (m/z)Product Ion 1 (m/z)Product Ion 2 (m/z)
1.Cyromazine1.3167.185.1125.1
2.Dinotefuran3.4203.1129.2157.2
3.Flonicamid3.6230.1203.1174.0
4.Nitenpyram3.7271.0225.2126.0
5.Pymetrozine3.8218.0105.078.0
6.Imidacloprid4.6256.0209.1175.1
7.Cymoxanil4.7198.9128.2111.2
8.Acetamiprid4.9223.0126.099.0
9.Carbetamide5.1237.1192.0118.1
10.Oxadixyl5.1279.1219.1132.1
11.Pyracarbolid5.5218.1125.097.0
12.Imazalil5.6297.0159.0201.0
13.Metalaxyl6.1280.1220.2192.2
14.Prometon6.1226.1142.086.0
15.Secbumeton6.1226.2170.1100.0
16.Terbumeton6.2226.1170.1100.0
17.Chlorantraniliprole6.4484.0452.9285.9
18.Ametryn6.6228.1186.196.0
19.Halofenozide6.6331.2105.0275.1
20.Furalaxyl6.7302.1242.195.0
21.Spiroxamine isomer 16.7298.2144.2100.1
22.Azoxystrobin6.9404.1372.1344.1
23.Fludioxinil6.9266.0227.1229.0
24.Spriroxamine isomer 26.9298.2144.2100.1
25.Flutolanil7.0324.1242.1262.1
26.Mandipropamid7.1412.1328.1356.1
27.Mepronil7.1270.1119.1228.0
28.Pyrimethanil7.1200.0107.082.0
29.Boscalid7.2343.0307.0140.0
30.Methoxyfenozide7.2369.1313.2149.1
31.Triadimefon7.2294.0197.1225.0
32.Fenhexamid7.4302.097.055.0
33.Myclobutanil7.5289.070.0125.0
34.Butafenacil7.6492.1331.0349.0
35.Fluoxastrobin8.1459.2427.2188.0
36.Mepanipyrim8.1224.0106.077.0
37.Tebufenozide8.2353.2133.0297.2
38.Picoxystrobin8.3368.0145.0205.0
39.Kresoxim-methyl8.4314.2116.1131.2
40.Carfentrazone-ethyl8.4412.0346.0366.0
41.Dimoxystrobin8.5327.1116.0205.0
42.Rotenone8.6395.1213.1192.1
43.Cyprodinil8.7226.093.077.0
44.Zoxamide8.7336.1187.0159.0
45.Amitraz8.9294.2148.391.2
46.Benalaxyl8.9326.2294.1148.1
47.Famoxadone9.0392.0331.0238.0
48.Benzoximate9.5364.0199.0105.0
49.Prochloraz9.7376.0308.070.0
50.Clofentezine9.8303.0138.0102.0
51.Trifloxystrobin9.9409.0186.0206.0
52.Fluazinam*10.3465.0148.991.0
53.Metaflumizone10.5507.1178.1287.1
54.Piperonyl butoxide10.5356.2177.2119.1
55.Tebufenpyrad10.5334.0145.0117.0
56.Pyriproxyfen10.6322.096.0185.0
57.Quinoxyfen10.9308.1197.1162.1
58.Avermectin B1b11.1876.5553.4291.0
59.Avermectin B1a11.3890.5567.5305.0
60.Fenpyroximate11.3422.0366.1135.1
61.Eprinomectin11.3914.6186.2154.2
62.Doramectin11.5916.6331.4593.5
63.Fenazaquin11.6307.1161.1147.0
64.Moxidectin11.6640.4528.5498.5
65.Ivermectin11.7892.6569.5307.3
*Fluazinam has improved response in negative mode (see inset). Isomers are numbered sequentially based on elution order. To download this retention table in XLS format, visit www.restek.com/LC-multiresidue
image
LC_FF0520
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:50 °C
SampleLC multiresidue pesticide standard #4 (cat.# 31975)
Diluent:Water
Conc.:50 ng/mL
Inj. Vol.:20 µL
Mobile Phase
A:Water + 4 mM ammonium formate + 0.1% formic acid
B:Methanol + 4 mM ammonium formate + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.59010
1.500.59010
4.000.54060
8.000.53070
11.000.50100
12.000.50100
12.010.59010
15.000.59010
Max Pressure:255 bar
DetectorAB SCIEX API 4000 LC-MS/MS
Ion Source:TurboIonSpray®
Curtain Gas:30 psi (206.8 kPa)
Gas 1:40 psi (275.8 kPa)
Gas 2:45 psi (310.3 kPa)
CAD:10 psi (68.9 kPa)
Source Temp.:350 °C
Mode:Scheduled MRM
MRM Detection Window:60 sec
Target Scan Time:0.33 sec
InstrumentAPI LC-MS/MS
NotesQ1/Q3 Resolution: Unit
Autosampler Temp.: 5 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

pesticides

,

multi-residue pesticides

,

multiresidue

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us