Amino Acids (9-Fluorenylmethyl-Chloroformate Derivatives) on Raptor ARC-18
This chromatogram may contain analytes that are synonyms of your search terms.

Amino Acids (9-Fluorenylmethyl-Chloroformate Derivatives) on Raptor ARC-18

PeakstR (min)Conc.
(µmole/mL)
1.Arginine (Arg)5.1450.4
2.Asparagine (Asn)5.5180.4
3.Glutamine (Gln)5.6300.4
4.trans-4-Hydroxyproline (Hyp)5.8780.4
5.Serine (Ser)6.1390.4
6.Aspartic acid (Asp)6.2560.4
7.Glutamic acid (Glu)6.6960.4
8.Threonine (Thr)7.0270.4
9.Glycine (Gly)7.4660.4
10.Alanine (Ala)8.2270.4
11.Proline (Pro)8.8030.4
12.Methionine (Met)9.3240.4
13.Valine (Val)9.5370.4
14.Norvaline9.6560.4
15.Tryptophan (Trp)9.9100.4
16.Phenylalanine (Phe)10.1200.4
17.Isoleucine (Ile)10.2140.4
18.Leucine (Leu)10.2850.4
19.Cystine10.7490.2
20.Histidine (His)12.0450.4
21.Lysine (Lys)12.1740.4
22.Cysteine (Cys)12.4630.4
23.Tyrosine (Tyr)12.8010.4
Amino acids are analyzed as 9-fluorenylmethyl-chloroformate derivatives.
image
LC_FF0539
ColumnRaptor ARC-18 (cat.# 9314A1E)
Dimensions:100 mm x 3 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Sample
Diluent:0.1 N HCl
Conc.: 0.4 μmole/mL for each amino acid (0.2 μmole/mL for cystine)
Inj. Vol.:1 µL
Mobile Phase
A:0.1% Formic acid + 20 mM ammonium formate in water
B:0.1% Formic acid + 10 mM ammonium formate in 90:10 acetonitrile:water
Time (min)Flow (mL/min)%A%B
0.000.88020
6.250.86040
9.000.84060
10.000.84060
13.000.80100
13.010.88020
15.000.88020
DetectorUV/Vis @ 265, 4.8 nm
InstrumentWaters ACQUITY UPLC H-Class
NotesDerivatization reaction: 50 μL amino acid mix + 100 μL 0.2 N borate buffer (pH 10.0) + 50 μL 15 mM 9-fluorenylmethyl-chloroformate solution + 50 μL acetonitrile

The injection can be performed after 5 minutes of reaction time.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

amino acid

,

FMOC

,

Raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us