share:
delim
delim
Resveratrol and Glucoside Derivative Isomers on Raptor ARC-18 by LC-MS/MS

Resveratrol and Glucoside Derivative Isomers on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.trans-Piceid*0.78389.1227.0142.9
2.cis-Piceid*0.94389.1227.0142.9
3.trans-Resveratrol1.06227.0184.9142.6
4.cis-Resveratrol1.17227.0184.9142.6
*Glucoside derivatives of trans-resveratrol and cis-resveratrol.
image
LC_FF0554
ColumnRaptor ARC-18 (cat.# 9314552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Temp.:40 °C
Sample
Diluent:Water
Conc.:25 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.1% formic acid
B:Acetonitrile + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.58515
1.500.53565
1.510.58515
2.500.58515
DetectorMS/MS
Ion Mode:ESI-
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

polydatin

,

piceid

,

antioxidant

,

polyphenol

,

phytoalexin

,

stilbenoid

,

glucoside

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us