share:
delim
delim
Anthocyanidins on Raptor ARC-18 5 μm by HPLC

Anthocyanidins on Raptor ARC-18 5 μm by HPLC

PeakstR (min)
1.Delphinidin0.724
2.Cyanidin0.992
3.Pelargonidin1.288
4.Malvidin1.442
image
LC_FF0556
ColumnRaptor ARC-18 (cat.# 931455E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:5 µm
Temp.:40 °C
Sample
Diluent:0.1% Trifluoroacetic acid in 50:50 acetonitrile:water
Conc.:30 µg/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Trifluoroacetic acid in water
B:0.1% Trifluoroacetic acid in methanol
Time (min)Flow (mL/min)%B
0.001.035
2.001.055
2.011.035
4.001.035
DetectorPDA @ 530 nm
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Anthocyanidins

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us