share:
delim
delim
Flavonols on Raptor C18 by LC-MS/MS

Flavonols on Raptor C18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion
1.Rutin1.04611.5303.3
2.Quercitrin1.19449.4303.3
3.Myricetin1.26319.3153.2
4.Fisetin1.27287.3137.3
5.Morin1.36303.3153.2
6.Quercetin1.48303.3153.2
7.Kaempferol1.68287.3153.2
8.Isorhamnetin1.72317.3153.2
9.Rhamnetin1.91317.3123.3
10.Galangin2.15271.3153.2
image
LC_FF0557
ColumnRaptor C18 (cat.# 9304552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:90:10 (water:acetonitrile)
Conc.:500 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.59010
2.400.53070
2.410.59010
3.500.59010
DetectorMS/MS
Ion Mode:ESI+
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

flavonol

,

raptor

,

raptor C18

,

C18

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us