share:
delim
delim
Water Soluble Vitamins by LC-MS/MS

Water Soluble Vitamins by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Thiamine (B1)1.2650265.0122.1144.1
2.Pyridoxine (B6)1.3520170.1152.1134.1
3.Niacin (B3)1.84200123.180.178.1
4.Pantothenic acid (B5)1.94200220.090.172.1
5.Folic acid (B9)2.78500442.1295.0176.3
6.Cyanocobalamin (B12)2.92500678.3147.2359.0
7.Biotin (B7)2.93100245.0227.197.1
8.Riboflavin (B2)3.12200377.0243.1172.1
image
LC_FF0558
ColumnUltra Aqueous C18 (cat.# 9178312)
Dimensions:100 mm x 2.1 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:40 °C
Sample
Diluent:Water
Inj. Vol.:2 µL
Mobile Phase
A:Water + 5 mM ammonium formate + 0.1% formic acid
B:Methanol + 5 mM ammonium formate + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.61000
3.500.63565
3.510.61000
6.000.61000
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesAutosampler temperature = 5 °C
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us