share:
delim
delim
Sulfonamide Antibiotics on Raptor C18 by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Sulfonamide Antibiotics on Raptor C18 by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Sulfanilamide0.83200172.9893.0775.23
2.Sulfadiazine1.4520251.18156.0492.08
3.Sulfathiazole1.6010256.16156.0392.08
4.Sulfapyridine1.6710250.13156.1092.08
5.Sulfamerazine1.7920265.08156.0392.08
6.Sulfamethazine2.0710279.23186.08124.08
7.Sulfamethizole2.1110271.17156.02108.02
8.Sulfamethoxypyridazine2.1610281.14156.03126.07
9.Sulfachlorpyridazine2.5520285.05156.03108.09
10.Sulfadoxine2.7510311.17156.03108.09
11.Sulfamethoxazole2.7820254.18155.98147.06
12.Sulfaethoxypyridazine2.8420295.17267.07156.03
13.Sulfisoxazole2.9820268.14156.03113.10
14.Sulfadimethoxine3.3710311.17156.09108.09
15.Sulfaquinoxaline3.4020301.18156.04108.02
image
LC_FS0501
ColumnRaptor C18 (cat.# 9304A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor C18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9304A0252)
Temp.:40 °C
Sample
Diluent:Water
Conc.:10–200 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.49010
3.500.46040
3.510.49010
5.000.49010
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Raptor

,

sulfonamide

,

antibiotic

,

residue

,

veterinary antibiotic

,

food producing animal

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us