share:
delim
delim
Morpholine Fungicides on Raptor C18

Morpholine Fungicides on Raptor C18

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Flumorph isomer 11.6220372.17285.09165.03
2.Flumorph isomer 21.7420372.17285.09165.03
3.Dimethomorph isomer 12.0020388.16301.08165.09
4.Dimethomorph isomer 22.1820388.16301.08165.09
5.Dodemorph isomer 12.8320282.27116.1298.11
6.Fenpropimorph3.8120304.27147.14130.16
7.Tridemorph isomer 14.0520298.33130.1798.11
8.Dodemorph isomer 24.3820282.27116.1298.11
9.Tridemorph isomer 24.4620298.33130.1798.11
image
LC_FS0520
ColumnRaptor C18 (cat.# 9304A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:50 °C
Sample
Diluent:Acetonitrile
Conc.:20 ng/mL
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium acetate, 0.1% acetic acid
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.45050
4.000.40100
4.500.40100
4.510.45050
5.500.45050
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

morpholine

,

morpholine fungicide

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us