share:
delim
delim
LC Multiresidue Pesticide Mix Extracted from Brown Rice Flour with QuEChERS Slim Pouch on Raptor ARC-18 by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

LC Multiresidue Pesticide Mix Extracted from Brown Rice Flour with QuEChERS Slim Pouch on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Cyromazine0.747167.168.143.1
2.Methamidophos1.318142.094.0125.1
3.Formetanate1.681222.1165.046.1
4.Carbofuran1.682222.2165.0123.1
5.Methabenzthiazuron1.686222.0165.0150.0
6.Acephate1.690184.0143.049.0
7.Pymetrozine1.708218.1105.078.0
8.Aminocarb1.716209.1137.1152.1
9.Propamocarb1.735189.1102.0144.1
10.Omethoate1.905214.0125.0183.0
11.Dinotefuran1.972203.1129.187.1
12.Aldicarb sulfoxide2.010207.189.0132.1
13.Butoxycarboxim2.022223.0106.0166.0
14.Nitenpyram2.049271.0225.056.0
15.Aldicarb sulfone2.062240.2148.086.0
16.Carbendazim2.090192.0160.0131.9
17.Oxamyl2.154237.172.090.0
18.Flonicamid2.234230.1203.0148.0
19.Methomyl2.259163.188.0106.1
20.Thiabendazole2.284201.892.2104.0
21.Thiamethoxam2.289292.1211.1181.1
22.Fuberidazole2.368185.0157.1103.0
23.Monocrotophos2.375224.1127.0193.1
24.Mexacarbate2.393223.2166.1151.1
25.Dioxacarb2.397224.1167.077.1
26.Dicrotophos2.526238.1112.172.0
27.Imidacloprid2.536256.1175.0209.0
28.Clothianidin2.545250.0169.0113.0
29.Dimethoate-d6 (IS)2.652236.2205.2131.2
30.Fenuron2.662165.072.146.1
31.Trichlorfon2.667257.0108.9220.8
32.Dimethoate2.668230.0199.0125.0
33.3-Hydroxycarbofuran2.679238.1181.0163.0
34.Vamidothion2.695288.0146.0118.0
35.Mevinphos2.715225.1127.0127.0
36.Acetamiprid2.724223.0126.056.1
37.Propoxur2.777210.0111.0168.0
38.Ethirimol2.780210.2140.298.2
39.Cymoxanil2.805199.1128.2111.2
40.Thiacloprid2.865253.0126.099.0
41.Pirimicarb2.896239.272.0182.1
42.Aldicarb3.022208.2116.289.2
43.Tricyclazole3.066190.0163.0136.0
44.Oxadixyl3.076279.1219.1132.1
45.Carbetamide3.085237.1192.0118.1
46.Simetryn3.122214.168.096.0
47.Thiophanate-methyl3.146343.0151.0311.1
48.Thidiazuron3.150221.1102.1127.9
49.Bendiocarb3.212224.1167.1109.0
50.Imazalil3.225297.1159.041.1
51.Prometon3.227226.1142.1184.0
52.Metribuzin3.230215.1187.1131.0
53.Secbumeton3.232226.2170.2100.2
54.Terbumeton3.235226.1114.1170.1
55.Sulfentrazone*3.263387.0309.0201.0
56.Pyracarbolid3.279218.1125.197.0
57.Tebuthiuron3.310229.1172.1116.0
58.Carbaryl3.316202.1145.0127.0
59.Carboxin3.326236.1143.087.1
60.Monolinuron3.373215.1126.0148.0
61.Fluometuron3.381233.172.146.1
62.Diuron3.382233.172.146.0
63.Ethiofencarb3.406226.0107.0164.0
64.Ametryn3.447228.168.1186.1
65.Chlorotoluron3.450213.072.146.1
66.Metobromuron3.483259.0170.0148.1
67.Propham3.500180.1120.277.1
68.Flutriafol3.508302.070.1123.0
69.Methoprotryne3.532272.1170.0198.1
70.Forchlorfenuron3.536248.1129.093.0
71.Isoprocarb3.537194.195.1137.1
72.Fenpropimorph3.561304.2147.1117.0
73.Atrazine-d5 (SS)3.575221.2179.2101.3
74.Isocarbophos3.606291.1231.1121.1
75.Isoproturon3.612207.172.146.1
76.Desmedipham3.620318.0154.1301.0
77.Phenmedipham3.623318.1108.193.1
78.Spiroxamine3.626298.2144.258.2
79.Pyrimethanil3.659200.0107.082.0
80.Metalaxyl3.691280.1220.1192.1
81.Chlorantraniliprole3.730484.0452.8285.8
82.Cycluron3.785199.189.146.0
83.Prometryn3.830242.1158.0200.2
84.Linuron-d6 (SS)3.843255.1160.1185.2
85.Linuron3.854249.0160.0182.0
86.Thiofanox3.876242.157.2184.0
87.Fenobucarb3.894208.095.1152.1
88.Azoxystrobin3.915404.2329.0344.0
89.Diethofencarb3.924268.1226.1124.0
90.Ethofumesate3.925304.1121.1241.0
91.Ethiprole3.932397.0350.9255.0
92.Fludioxonil*3.935247.1180.1126.1
93.Methiocarb3.958226.1169.1121.1
94.Siduron3.967233.394.0137.1
95.Halofenozide3.974331.1104.9275.0
96.Fenamidone3.993312.092.1236.1
97.Furalaxyl3.995302.1242.195.0
98.Acibenzolar-S-methyl4.023210.9136.191.1
99.Nuarimol4.030315.1252.0207.0
100.Dimethomorph4.044388.2300.9165.1
101.Boscalid4.050343.0306.8140.0
102.Mandipropamid4.067412.1328.1125.0
103.Flutolanil4.079324.1262.0242.0
104.Promecarb4.104208.1109.0151.0
105.Paclobutrazol4.119294.070.1125.0
106.Triadimenol4.119296.170.143.1
107.Mepronil4.161270.191.1119.0
108.Cyproconazole4.175292.270.1125.0
109.Myclobutanil4.195289.270.1125.1
110.Chloroxuron4.225291.172.146.1
111.Bupirimate4.226317.2166.0237.1
112.Methoxyfenozide4.230369.2149.0313.1
113.Triadimefon4.287294.069.0197.0
114.Bifenazate4.290301.1198.0170.1
115.Cyprodinil4.313226.193.077.0
116.Mefenacet4.325299.1148.1120.1
117.Mepanipyrim4.334224.177.1106.1
118.Fluquinconazole4.346376.0348.9307.0
119.Fenhexamid4.383302.197.155.1
120.Fluoxastrobin4.385459.2427.0188.1
121.Tetraconazole4.387372.0159.070.0
122.Bromuconazole isomer 14.392376.0159.0159.0
123.Butafenacil4.411492.1330.9179.9
124.Iprovalicarb4.412321.2119.091.1
125.Flufenacet4.427364.1152.0194.1
126.Triticonazole4.429318.070.0125.0
127.Cyazofamid4.441325.1108.044.0
128.Diflubenzuron4.464311.0158.0141.0
129.Spirotetramat4.478374.1330.1302.0
130.Fenarimol4.510331.0267.9189.1
131.Fenbuconazole4.516337.0125.070.1
132.Epoxiconazole4.529330.0121.1101.0
133.Etaconazole4.552328.1159.055.0
134.Fipronil*4.571436.9332.0252.0
135.Neburon4.635275.057.188.1
136.Flusilazole4.646316.1247.1165.1
137.Fenoxycarb4.653302.288.1116.1
138.Picoxystrobin4.674368.0145.0205.0
139.Rotenone4.683395.1213.1192.1
140.Tebufenozide4.719353.2133.1105.0
141.Dimoxystrobin4.738327.1205.1116.1
142.Bromuconazole isomer 24.740378.0159.0159.0
143.Flubendiamide4.745700.0408.0274.0
144.Diclobutrazol4.764328.070.1159.0
145.Carfentrazone ethyl4.773429.1412.1346.1
146.Kresoxim-methyl4.795314.0267.1116.0
147.Tebuconazole4.827308.070.1124.9
148.Spinosad A4.840732.5142.198.1
149.Penconazole4.853284.170.1159.0
150.Prothioconazole4.857344.0189.0154.0
151.Zoxamide4.893336.1187.0159.0
152.Famoxadone4.933392.2331.1238.0
153.Diazinon-diethyl-d10 (IS)4.961315.2170.2154.3
154.Hydramethylnon4.979495.2323.245.2
155.Triflumuron4.986359.1156.0111.0
156.Prochloraz4.986376.0308.070.1
157.Hexaconazole4.993314.170.0158.8
158.Benalaxyl5.021326.2148.0208.1
159.Metconazole5.033320.170.1125.0
160.Propiconazole5.054342.0159.069.0
161.Pyraclostrobin5.076388.0194.0163.1
162.Clofentezine5.080303.0138.0102.0
163.Bitertanol5.120338.2269.270.1
164.Spinosad D5.142746.5142.098.0
165.Benzoximate5.171364.0199.1105.0
166.Diniconazole5.184326.170.0158.9
167.Thiobencarb5.197258.0125.089.1
168.Pencycuron5.229329.1125.089.0
169.Spinetoram5.234748.5142.198.2
170.Hexaflumuron*5.351459.0439.0175.0
171.Triflumizole5.366346.1278.073.1
172.Indoxacarb5.375528.1149.9203.0
173.Difenoconazole5.380406.1251.1152.0
174.Ipconazole5.426334.270.0124.9
175.Trifloxystrobin5.426409.0186.0206.0
176.Novaluron*5.455491.0471.0304.9
177.Emamectin-benzoate B1b5.463872.2158.282.1
178.Clethodim5.565360.1164.1268.1
179.Teflubenzuron5.614380.9140.9158.0
180.Alanycarb5.636400.191.0238.0
181.Buprofezin5.642306.2106.186.0
182.Emamectin-benzoate B1a5.644886.5158.082.1
183.Benfuracarb5.695411.2194.9251.9
184.Fluazinam*5.726463.0415.9397.9
185.Metaflumizone*5.728505.1302.1328.0
186.Furathiocarb5.735383.1195.0252.1
187.Temephos5.757467.1125.0418.9
188.Lufenuron5.767511.2141.0158.0
189.Tebufenpyrad5.778334.0117.0145.0
190.Pyriproxyfen5.843322.096.0227.1
191.Piperonyl butoxide5.864356.2177.0119.1
192.Hexythiazox5.924353.2228.0168.1
193.Quinoxyfen5.935308.1197.0162.0
194.Flufenoxuron5.984489.1158.0141.0
195.Propargite6.088368.1231.1175.0
196.Etoxazole6.099360.2141.057.1
197.Spiromesifen6.129371.3273.2255.2
198.Chlorfluazuron6.165539.8382.9158.0
199.Spirodiclofen6.269428.371.1313.1
200.Fenpyroximate6.283422.1366.1138.1
201.Pyridaben6.442365.2147.0309.1
202.Eprinomectin6.466914.6186.1154.1
203.Abamectin B1a6.544890.6305.4567.2
204.Fenazaquin6.559307.1161.0131.1
205.Doramectin6.733916.6331.2593.4
206.Ivermectin6.893892.6551.4569.4
*Detected in ESI- mode
image
LC_FS0521
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9314A0252)
Temp.:50 °C
SampleLC multiresidue pesticide kit (cat.# 31971)
Diluent:Water:acetonitrile 9:1
Conc.:1 ng/mL see notes for details
Inj. Vol.:5 µL
Mobile Phase
A:0.2% Formic acid, 2 mM ammonium formate in water
B:0.2% Formic acid, 2 mM ammonium formate in methanol
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.500.40100
7.510.4955
9.500.4955
DetectorMS/MS
Ion Mode:ESI+/ESI-
Mode:Scheduled MRM
InstrumentUHPLC
NotesAlthough the ten LC multiresidue pesticides kit mixes are formulated to ensure maximum long-term stability and reliability as packaged, stability may become an issue when a large number of compounds with different chemical functionalities are combined together into a single mix. This should be taken into consideration for quantitative analysis.

Sample Preparation:
Brown rice flour (10 g), purchased from a local market, was mixed with 10 mL water. After mixing for 10 min, the brown rice flour was fortified at 10 ng/g with all residues from the LC multiresidue pesticide kit (cat.# 31971), which had been previously combined into a single mixture that day. The fortified brown rice flour suspension was then mixed for 30 min to ensure uniform dispersion of the added pesticide residues in the brown rice flour suspension. Extraction was performed using 10 mL of acetonitrile and original unbuffered QuEChERS salts (cat.# 25848). After centrifugation, 1 mL supernatant was added to a 2 mL dSPE vial containing magnesium sulfate, PSA, and C18 (cat.# 26125) for cleanup. After centrifugation, 100 μL of the 10 ng/mL extract was diluted with 900 μL water and injected onto the UHPLC.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Slim Pouch

,

Brown

,

Rice

,

Flour

,

QuEChERS

,

Pesticides

,

dSPE

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us