Mycotoxins in Peanut Powder on Raptor Biphenyl by LC-MS/MS
This chromatogram may contain analytes that are synonyms of your search terms.

Mycotoxins in Peanut Powder on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Conc.
(ng/g)
Precursor IonProduct Ion 1Product Ion 2
1.Deoxynivalenol0.6250297.3249.3231.2
2.Fumonisin B12.4550722.5352.4334.5
3.HT-22.6050447.3345.3285.3
4.Fumonisin B32.8550706.5336.4318.4
5.Fumonisin B23.2350706.5336.3141.2
6.T23.3150489.3245.2387.4
7.Aflatoxin G23.745331.2313.3189.3
8.Zearalenone3.9650319.3283.3187.2
9.Aflatoxin G14.225329.2243.2200.2
10.Aflatoxin B24.435315.3287.3259.2
11.Aflatoxin B14.995313.3285.2241.2
12.Ochratoxin A5.195404.2239.3358.3
Matrix interference was observed for HT-2 transition 447.3-285.3 in peanut powder. Therefore, 447.3-345.3 was chose for quantification.
image
LC_FS0526
ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:40 °C
Inj. Vol.:5 µL
Mobile Phase
A:Water, 2 mM ammonium formate, 0.1% formic acid
B:Methanol, 2 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B%C
0.000.57030
0.60.57030
0.70.55050
3.00.52570
4.50.52575
5.00.51090
5.20.51090
5.210.52575
6.000.52575
6.010.57030
7.000.57030
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesWeighed 1.00 gram of peanut powder in a 50 mL centrifuge tube and added 2.00 mL of water. Vortexed at 3000 rpm for 5 min followed by the addition of 4.0 mL of extraction solvent (50:50 water:acetonitrile, v/v). The tube was then vortexed at 3000 rpm for 5 min followed by centrifugation for 15 min at 4200 rpm. 475 μL of the supernatant was filtered through a Thomson SINGLE StEP Nano filter vial (0.2 μm, cat.# 25882). The sample was then fortified with 25 μL of a standard solution prepared in water at 1000 ng/mL (100 ng/mL for aflatoxins and ochratoxin A) as part of the matrix-matched calibration curve. Vortexed at 3000 rpm for 1 min prior to analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Mycotoxins

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us