Polar Pesticides in Water at 20 ppt on Raptor Polar X by LC-MS/MS

Polar Pesticides in Water at 20 ppt on Raptor Polar X by LC-MS/MS

PeakstR (min)Conc.
(ng/L)
Precursor IonProduct Ion 1Product Ion 2Product Ion 3
1.AMPA1.11220110.0078.9062.9081.00
2.Glufosinate4.91320180.2094.9563.0085.00
3.Glyphosate5.95920168.1062.9578.95-
image
LC_FS0549
ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Diluent:HPLC water
Inj. Vol.:500 µL
Mobile Phase
A:Water, 0.5% formic acid
B:Acetonitrile, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.5595
1.00.5595
3.00.59010
9.00.59010
9.010.5595
10.50.5595
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
Download PDF