share:
delim
delim
16 Cannabinoids on Raptor ARC-18 1.8 μm by LC-UV
This chromatogram may contain analytes that are synonyms of your search terms.

16 Cannabinoids on Raptor ARC-18 1.8 μm by LC-UV

PeakstR (min)
1.Cannabidivarinic acid (CBDVA)0.787
2.Cannabidivarin (CBDV)0.867
3.Cannabidiolic acid (CBDA)1.090
4.Cannabigerolic acid (CBGA)1.157
5.Cannabigerol (CBG)1.216
6.Cannabidiol (CBD)1.279
7.Tetrahydrocannabivarin (THCV)1.396
8.Tetrahydrocannabivarinic acid (THCVA)1.774
PeakstR (min)
9.Cannabinol (CBN)1.894
10.Cannabinolic acid (CBNA)2.257
11.Δ9-Tetrahydrocannabinol (Δ9-THC)2.380
12.Δ8-Tetrahydrocannabinol (Δ8-THC)2.456
13.Cannabicyclol (CBL)2.826
14.Cannabichromene (CBC)2.966
15.Tetrahydrocannabinolic acid A (THCA-A)3.150
16.Cannabichromenic acid (CBCA)3.516
image
LC_GN0579
ColumnRaptor ARC-18 (cat.# 931421E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:1.8 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter, 0.2 μm frit (cat.# 25809)
Temp.:30 °C
SampleTetrahydrocannabivarin (cat.# 34100)
Cannabidiolic acid (CBDA) (cat.# 34099)
Cannabichromene (CBC) (cat.# 34092)
Cannabigerol (CBG) (cat.# 34091)
delta-9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093)
delta-8-Tetrahydrocannabinol (Δ8-THC) (cat.# 34090)
delta-9-Tetrahydrocannabinol (Δ9-THC) (cat.# 34067)
Cannabinol (CBN) (cat.# 34010)
Cannabidiol (CBD) (cat.# 34011)
Compounds not present in these mixes were obtained separately.
Diluent:25:75 Water:methanol
Conc.:50 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.02575
4.001.02575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cannabinoids

,

cannabis

,

potency

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us