share:
delim
delim
Potency Analysis of a Commercially Available CBD Product on Raptor ARC-18 2.7 μm by LC-UV

Potency Analysis of a Commercially Available CBD Product on Raptor ARC-18 2.7 μm by LC-UV

PeakstR (min)
1.Cannabidiolic acid (CBDA)2.608
2.Cannabigerol (CBG)2.934
PeakstR (min)
3.Cannabidiol (CBD)3.101
4.Δ9-Tetrahydrocannabinol (Δ9-THC)5.851
5.Cannabichromene (CBC)7.315
image
LC_GN0581
ColumnRaptor ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.# 9314A0250)
Temp.:30 °C
Sample
Diluent:25:75 Water:methanol
Conc.: 6.24 mg/mL (CBD)
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.52575
9.001.52575
DetectorUV/Vis @ 228 nm
InstrumentHPLC
Notes50 μL of a commerically available CBD product was diluted in 950 μL of 25:75 water:methanol. The sample was vortexed for 30 seconds at 3000 rpm. 400 μL of the sample was then filtered using a Thomson SINGLE StEP standard filter vial (PVDF, 0.2 μm, Restek cat.# 25895) prior to analysis. CBDA, CBG, CBD, Δ9-THC, and CBC were detected, but only CBD was quantified.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cannabinoids

,

cannabis

,

potency

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us