share:
delim
delim
Cannabis Flower Sample on Raptor ARC-18 2.7 μm by HPLC-UV

Cannabis Flower Sample on Raptor ARC-18 2.7 μm by HPLC-UV

PeakstR (min)
1.Cannabidivarinic acid (CBDVA)2.72
2.Cannabigerolic acid (CBGA)2.90
PeakstR (min)
3.Tetrahydrocannabivarinic acid (THCVA)4.45
4.Δ9-Tetrahydrocannabinol (Δ9-THC)6.05
5.Tetrahydrocannabinolic acid A (THCA-A)7.87
image
LC_GN0588
ColumnRaptor ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.# 9314A0250)
Temp.:30 °C
Sample
Diluent:25:75 Water:methanol
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.52575
9.001.52575
DetectorUV/Vis @ 228 nm
InstrumentHPLC
Notes200 mg of ground flower was weighed into a 50 mL centrifuge tube. 10 mL of IPA was added prior to vortexing (5 seconds) and sonicating (5 minutes) over a total of three cycles. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 10-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cannabinoids

,

cannabis

,

potency

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us