Standard Analysis of 21 Cannabinoids on Raptor ARC-18 2.7 μm by LC-UV

Standard Analysis of 21 Cannabinoids on Raptor ARC-18 2.7 μm by LC-UV

PeakstR (min)
1.Cannabidivarinic acid (CBDVA)1.775
2.Cannabidivarin (CBDV)1.973
3.Cannabidiolic acid (CBDA)2.442
4.Cannabigerolic acid (CBGA)2.575
5.Cannabigerol (CBG)2.732
6.Cannabidiol (CBD)2.904
7.Tetrahydrocannabivarin (THCV)3.203
8.Cannabichromevarin (CBCV)3.867
9.Tetrahydrocannabivarinic acid (THCVA)4.023
10.Cannabinol (CBN)4.321
PeakstR (min)
11.Cannabidiphorol (CBDP)4.866
12.Cannabinolic acid (CBNA)5.099
13.Δ9-Tetrahydrocannabinol (Δ9-THC)5.460
14.Δ8-Tetrahydrocannabinol (Δ8-THC)5.637
15.Cannabicyclol (CBL)6.496
16.Cannabichromene (CBC)6.788
17.Tetrahydrocannabinolic acid A (THCA-A)7.122
18.Cannabichromenic acid (CBCA)7.926
19.Cannabicyclolic acid (CBLA)8.458
20.Tetrahydrocannabiphorol (THCP)10.251
21.Cannabicitran (CBT)11.015
image
LC_GN0668
ColumnRaptor ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9314A0250)
Temp.:30 °C
SampleCannabidivarinic acid (CBDVA) (cat.# 34134)
Cannabidivarin (CBDV) (cat.# 34123)
Cannabidiolic acid (CBDA) (cat.# 34094)
Cannabigerolic acid (CBGA) (cat.# 34135)
Cannabigerol (CBG) (cat.# 34091)
Cannabidiol (CBD) (cat.# 34011)
Tetrahydrocannabivarin (THCV) (cat.# 34100)
Cannabinol (CBN) (cat.# 34010)
d9-Tetrahydrocannabinol (d9-THC) (cat.# 34067)
d8-Tetrahydrocannabinol (d8-THC) (cat.# 34090)
Cannabicyclol (CBL) (cat.# 34130)
Cannabichromene (CBC) (cat.# 34092)
d9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34111)
Compounds not present in these mixes were obtained separately.
Diluent:Methanol
Conc.:50 ppm
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.52575
121.52575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us