Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

PFAS in Air, Part 1: OTM-45

22 September 2021
By
  • Jason Hoisington
Tags
  • #PFAS
  • #Product Selection
  • #Air
  • #Method Optimization
  • #Method Development
  • #Air & Gas Sampling Products
  • #Sample Preparation & Air Sampling
  • #Volatile Organic Compound (VOC) Sampling
  • #Blogs
  • #Environmental & Industrial Exposure
  • #Air
  • #Air Sampling
  • #LC
  • #MS/MS
  • Share:

If you’ve been following the news about PFAS, you’ll know they’re everywhere. They’re in the water, they’re in your clothes,  and now they’re in your air lab. Not in the sense that your lab has some PFAS background (although that’s likely true as well, see our PFAS method, product, and contaminant-free workflow for EPA Method 537.1 guides for more info), but earlier this year the EPA published OTM-45, a method covering the sampling and analysis of stack gasses for PFAS compounds.

https://www.epa.gov/sites/default/files/2021-01/documents/otm_45_semivolatile_pfas_1-13-21.pdf

If you’re not familiar with OTM (Other Test Methods), they are methods published which have not been subject to the federal rulemaking process. Per the EPA, “OTM 45 is a draft method under evaluation that will be updated as more data from stakeholders becomes available. Due to the urgent need for consistency, this method is being released as an “Other Test Method (OTM)” by EPA’s Emission Measurements Center to promote consistency with what we believe is the current best practices to sample and analyze the PFAS targets from stationary sources. We solicit any and all feedback, comments, and additional data coming from the application of this method as we work to adjust this method in anticipation of developing a reference method for PFAS from air emission sources.” In short, this is a draft method that the EPA hopes to get feedback on before finalizing, so if your lab is doing or considering PFAS work for stationary sources you may want to consider kicking the tires on OTM-45 so you can have input on the final version.

So what does OTM-45 entail? On the sample collection side, if you’re familiar with SW-846 0010 the sampling train will likely look familiar (see Fig. 1). Following that the various sections of the sampling train are extracted, resulting in 4 final samples. The front half rinse and filter (sections in red and yellow) are combined and extracted via shakeout. The back half rinse and first XAD-2 (green and blue) are combined and also extracted by shakeout. The condensate, impinger water, and impinger rinse (purple and grey) are combined and extracted by Weak Anion Exchange (WAX) SPE,  similar to EPA method 533.  Finally, the breakthrough XAD-2 (pink) is extracted via shakeout.

blog-PFAS-in-air-part-1-OTM-45-01.png

Fig. 1 – OTM-45 sample train

After extraction the samples are analyzed by LC-MS/MS using an isotope dilution calibration similar to EPA Method 533, though with a much broader compound list. With a lack of data showing what PFAS compounds are likely to be present the EPA seems to have taken the approach of casting the widest net possible. The number of compounds and isotope analogs for pre- and post-extraction internal standards unfortunately means that implementing OTM-45 is more difficult than just throwing new samples at your existing 533 LC-MS/MS method.

As  you’ve probably been spoiled by the “Part 1” in the blog title, I’ll be covering OTM-45 in multiple blogs, breaking things down to cover the LC-MS/MS analysis, calibration, sample prep, and background and cleanliness issues. There’s a lot to cover, so stay tuned.

References

  1. Method Guide for PFAS Analysis (https://www.restek.com/en/technical-literature-library/articles/method-guide-for-pfas-analysis)
  2. Product Guide for PFAS Analysis (https://www.restek.com/en/technical-literature-library/articles/product-guide-for-pfas-analysis/)
  3. Meet Requirements of EPA Method 537.1 PFAS Analysis with Contaminant-Free Workflow (https://www.restek.com/en/technical-literature-library/articles/meet-requirements-of-EPA-method-537.1-pfas-analysis-with-contaminant-free-workflow/)
  4. EPA OTM-45 (https://www.epa.gov/sites/default/files/2021-01/documents/otm_45_semivolatile_pfas_1-13-21.pdf)
  5. EPA Method 533 (https://www.epa.gov/sites/default/files/2019-12/documents/method-533-815b19020.pdf)

Related Videos

PFAS Analysis – Why a Delay Column is Important

View Video

Related Articles

Method Guide for PFAS Analysis

Read Article

Product Guide for PFAS Analysis

Read Article

Meet Requirements of EPA Method 537.1 PFAS Analysis with Contaminant-Free Workflow

Read Article

Related Products

Resprep Polymeric SPE Cartridges and 96-Well Plates (Reversed-Phase mode, Ion-Exchange mode)

View Product