Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

TO-15 65 Component Mix on Rxi-5ms (30 m)

GC_AR1149
PeakstR (min)
1.Propylene1.35
2.Dichlorodifluoromethane (Freon® 12)1.37
3.Chloromethane1.44
4.1,2-Dichlorotetrafluoroethane (Freon® 114)1.46
5.Vinyl chloride1.51
6.1,3-Butadiene1.55
7.Bromomethane1.68
8.Chloroethane1.74
9.Ethanol1.86
10.Trichlorofluoromethane (Freon® 11)1.95
11.Acrolein1.98
12.Acetone2.03
13.Acetonitrile (contaminant)2.04
14.Isopropyl alcohol2.13
15.1,1-Dichloroethene2.21
16.1,1,2-Trichlorotrifluoroethane (Freon® 113)2.29
17.Methylene chloride2.33
18.Carbon disulfide2.38
19.trans-1,2-Dichloroethene2.63
20.Methyl tert-butyl ether (MTBE)2.70
21.1,1-Dichloroethane2.78
22.Vinyl acetate2.84
23.2-Butanone (MEK)*3.01
PeakstR (min)
24.Hexane*3.01
25.cis-1,2-Dichloroethene3.14
26.Ethyl acetate3.24
27.Bromochloromethane (IS)3.27
28.Chloroform3.28
29.Tetrahydrofuran3.43
30.1,1,1-Trichloroethane3.67
31.1,2-Dichloroethane3.71
32.Benzene3.90
33.Carbon tetrachloride3.90
34.Cyclohexane3.91
35.1,4-Difluorobenzene (IS)4.15
36.Heptane4.44
37.1,2-Dichloropropane4.46
38.Trichloroethylene4.48
39.Bromodichloromethane4.59
40.Methyl methacrylate4.67
41.1,4-Dioxane4.75
42.4-Methyl-2-pentanone (MIBK)5.13
43.cis-1,3-Dichloropropene5.13
44.trans-1,3-Dichloropropene5.55
45.Toluene5.62
46.1,1,2-Trichloroethane5.67
PeakstR (min)
47.2-Hexanone (MBK)5.96
48.Dibromochloromethane6.11
49.1,2-Dibromoethane6.31
50.Tetrachloroethene6.37
51.Chlorobenzene-d5 (IS)6.97
52.Chlorobenzene7.01
53.Ethylbenzene7.25
54.m-Xylene7.39
55.p-Xylene7.39
56.Bromoform7.69
57.Styrene7.76
58.o-Xylene7.80
59.1,1,2,2-Tetrachloroethane8.09
60.4-Bromofluorobenzene**8.32
61.4-Ethyltoluene8.96
62.1,3,5-Trimethylbenzene9.04
63.1,2,4-Trimethylbenzene9.46
64.1,3-Dichlorobenzene9.69
65.Benzyl chloride9.79
66.1,4-Dichlorobenzene9.79
67.1,2-Dichlorobenzene10.17
68.1,2,4-Trichlorobenzene12.26
69.Naphthalene12.38
70.Hexachlorobutadiene12.66
*Peaks 23 and 24 share ion m/z 43; **Tuning standard
ColumnRxi-5ms, 30 m, 0.32 mm ID, 1.00 µm (cat.# 13454)
SampleTO-15 65 component mix (cat.# 34436)
TO-14A internal standard/tuning mix (cat.# 34408)
Diluent:Nitrogen
Conc.:10.0 ppbv 200 cc injection
InjectionDirect
Oven
Oven Temp.:32 °C (hold 1 min) to 150 °C at 11 °C/min to 230 °C at 33 °C/min
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
Linear Velocity:51 cm/sec @ 32 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:230 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:69.9 eV
Solvent Delay Time:1.0 min
Tune Type:BFB
Ionization Mode:EI
Scan Range:35-250 amu
Scan Rate:3.32 scans/sec
PreconcentratorNutech 8900DS
  Trap 1 Settings
Type/Sorbent :Glass beads
Cooling temp:-155 °C
Preheat temp:5 °C
Preheat time:0 sec
Desorb temp:20 °C
Desorb flow:5 mL/min
Desorb time:360 sec
Bakeout temp:200 °C
Flush flow:120 mL/min
Flush time:60 sec
Sweep flow:120 mL/min
Sweep time:60 sec
  Trap 2 Settings
Type/Sorbent:Tenax
Cooling temp:-35 °C
Desorb temp:190 °C
Desorb time:30 sec
Bakeout temp:200 °C
Bakeout time:10 sec
  Cryofocuser
Cooling temp:-160 °C
Inject time:140 sec
  Internal Standard
Purge flow:100 mL/min
Purge time:6 sec
Vol.:20 mL
ISTD flow:100 mL/min
  Standard
Size:200 mL
Purge flow:100 mL/min
Purge time:6 sec
Sample flow:100 mL/min
InstrumentHP6890 GC & 5973 MSD
AcknowledgementNutech