Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Derivatized Amphetamines (500 ng/mL) on Rxi-5Sil MS

GC_CF1144
Peaks
1.Amphetamine
2.Methamphetamine
3.MDA
4.MDMA
5.MDEA
c = contaminant
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Butyl chloride
Conc.:500 ng/mL HFAA derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5 mm splitless taper w/wool (cat.# 22286-200.1)
Inj. Temp.:250 °C
Purge Flow:28 mL/min
Oven
Oven Temp.:75 °C to 300 °C at 15 °C/min
Carrier GasHe, constant linear velocity
Linear Velocity:45 cm/sec, 13.5 psi, 93.1 kPa @ 75 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:4 min
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:40-300 amu
Scan Rate:5 scans/sec
InstrumentShimadzu 2010 GC & QP2010+ MS