Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Derivatized Amphetamines (50 ng/mL) on Rxi-5Sil MS

GC_CF1147
Peaksm/z 1m/z 2m/z 3
1.Amphetamine11824091
2.Methamphetamine25411891
3.MDA135162240
4.MDMA254162135
5.MDEA268240162
c = contaminant
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Butyl chloride
Conc.:50 ng/mL HFAA derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5mm Gooseneck Splitless w/Wool (cat.# 22286-200.1)
Inj. Temp.:250 °C
Purge Flow:28.8 mL/min
Oven
Oven Temp.:75 °C to 300 °C at 15 °C/min
Carrier GasHe, constant linear velocity
Linear Velocity:45 cm/sec, 13.5 psi, 93.1 kPa @ 75 °C
DetectorMS
Mode:SIM
SIM Program:91, 118, 240, 254, 135, 162, 268 m/z
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Solvent Delay Time:4 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS