Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Blood Alcohol Analysis on Rtx-BAC Plus 1

GC_CF1153
PeakstR (min)Conc.
(µg/mL)
1.Methanol0.912100
2.Acetaldehyde0.967100
3.Ethanol1.158100
4.Isopropanol1.420100
5.Acetone1.548100
6.tert-Butanol1.68850
7.1-Propanol (n-Propanol)1.96950
ColumnRtx-BAC Plus 1, 30 m, 0.32 mm ID, 1.8 µm (cat.# 18004)
SampleBAC resolution control standard n-P (cat.# 36010)
BAC resolution control standard t-B (cat.# 36011)
Diluent:Water
Conc.: 50 µL of each standard were diluted in 900 µL water in a 20 mL headspace vial.
Injectionheadspace-loop split (split ratio 50:1)
Liner:1 mm ID straight inlet liner (cat.# 20972)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:3 min
Transfer Line Temp.:125 °C
Valve Oven Temp.:125 °C
Standby flow rate:50 mL/min
Sample Temp.:60 °C
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:1 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 3 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec @ 40 °C
DetectorFID @ 240 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesHeadspace concentrator courtesy of Teledyne Tekmar, Mason, OH.