Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Inhalants on Rtx-BAC Plus 1 and Plus 2 Column Set

GC_CF1162
PeaksRtx-BAC Plus 1 tR (min)Rtx-BAC Plus 2 tR (min)
1.Methanol1.3401.230
2.Ethanol1.6941.470
3.Isopropanol2.0721.660
4.Diethyl ether2.1491.334
5.Acetone2.2561.589
6.Methylene chloride2.4051.703
7.MTBE3.1851.660
8.Hexane3.6831.660
9.Chloroform3.9562.667
PeaksRtx-BAC Plus 1 tR (min)Rtx-BAC Plus 2 tR (min)
10.MEK4.2102.543
11.Ethyl acetate4.5242.414
12.Benzene5.3902.883
13.Trichloroethylene6.3063.522
14.MIBK8.1015.775
15.Toluene8.1625.440
16.Ethylbenzene10.2867.681
17.m-Xylene10.4877.879
18.p-Xylene10.4877.879
19.o-Xylene10.9908.467
ColumnsRtx-BAC Plus 1 30 m, 0.32 mm ID, 1.8 µm (cat.# 18004)
and Rtx-BAC Plus 2 30 m, 0.32 mm ID, 0.6 µm (cat.# 18006)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal “Y” Press-Tight connector (cat.# 20405-261)
Standard/Sample
Diluent:Water
Conc.:50 µg/mL each inhalant
Injectionheadspace-loop split (split ratio 50:1)
Liner:1 mm straight inlet liner (cat.# 20972)
Headspace-Loop
Inj. Port Temp.:220 °C
Instrument:Tekmar HT3
Inj. Time:3 min
Transfer Line Temp.:125 °C
Valve Oven Temp.:125 °C
Sample Temp.:70 °C
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 4 min) to 120 °C at 10 °C/min (hold 0 min)
Carrier GasHe, constant flow
Linear Velocity:50 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesThe Rtx-BAC Plus 1 and Plus 2 columns were connected to the injection port using a ~12 inch section of guard column between the injection port and the universal Y Press-Tight connector.

Headspace concentrator courtesy of Teledyne Tekmar, Mason, OH.