Amine Test Mix Rtx-35 Amine

ColumnRtx-35 Amine, 30 m, 0.53 mm ID, 1.00 µm (cat.# 11355)
SampleAmine column test mix (cat.# 35002)
Diluent:Methanol/methylene chloride
Conc.:450-900 ppm
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Splitless taper (4 mm), base deactivated (cat.# 20798-210.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:110 °C (hold 4 min) to 200 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec
DetectorFID @ 300 °C