Phenols & Benzidines US EPA Methods 604 & 605 on Rtx-5

ColumnRtx-5, 30 m, 0.53 mm ID, 1.50 µm (cat.# 10270)
Sample604 Phenols Calibration Mix (cat.# 31029)
605 Benzidines Calibration Mix (cat.# 31030)
Conc.: concentration approximately 25ng per component
Injection
Inj. Vol.:2.5 µL splitless
Inj. Temp.:300 °C
Oven
Oven Temp.:40 °C (hold 6 min) to 300 °C at 10 °C/min (hold 15 min)
Carrier GasH2, constant linear velocity
Linear Velocity:80 cm/sec
DetectorFID @ 300 °C
NotesFlow rate: 10 cc/min