Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides (0.32mm ID)

ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139)
SampleOrganochlorine Pesticide Mix AB #2 (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000)
Conc.:16-160 pg
InjectionDirect
Liner:Unliner liner (cat.# 20335)
Inj. Temp.:200 °C
Oven
Oven Temp.:120 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.3 mL/min @ 120 °C
Dead Time:1.9 min
DetectorECD @ 300 °C
NotesECD with anode purge
Head pressure: 8.7psi (constant)

See Also: Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides2 (0.32mm ID) GC_EV00075A