Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides2 (0.32mm ID)

ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
SampleOrganochlorine Pesticide Mix AB #2 (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000)
Conc.:16-160 pg
InjectionDirect
Liner:Uniliner liner (cat.# 20335)
Oven
Oven Temp.:120 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.3 mL/min @ 120 °C
Dead Time:1.9 min
DetectorECD @ 300 °C
NotesECD with anode purge
Head pressure: 8.7psi (constant)

See also: Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides (0.32mm ID) GC_EV00075