Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Chlorophenoxyacid Herbicides (methyl esters) US EPA Method 515 by GC/ECD on Rtx-CLPesticides2

ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
Sample515.4 Methylated Chlorinated Acids Mix (cat.# 32444)
Diluent:MTBE
Conc.: 10-100µg/mL each methyl ester
Injection
Inj. Vol.:1.0 µL splitless (hold 0.45 min)
Liner:Double Gooseneck Splitless (4mm) (cat.# 20784)
Inj. Temp.:225 °C
Oven
Oven Temp.:70 °C (hold 1 min) to 210 °C at 20 °C/min (hold 1 min) to 300 °C at 5 °C/min
Carrier GasHe, constant pressure (10.5 psi, 72.4 kPa)
Temp.:70 °C
DetectorECD @ 320 °C