Chlorinated Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides by EPA 551.1 on Rtx-1301

ColumnRtx-1301, 30 m, 0.25 mm ID, 1.00 µm (cat.# 16053)
SampleDisinfection Byproducts and Chlorinated Solvents Mix (cat.# 30615)
Method 551.1 Pesticide/Herbicide Mix (cat.# 32438)
Conc.:5-10 µg/mL each analyte
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:4mm Split w/Wool (cat.# 20781)
Inj. Temp.:200 °C
Oven
Oven Temp.:35 °C (hold 22 min) to 145 °C at 10 °C/min (hold 2 min) to 225 °C at 20 °C/min (hold 15 min) to 260 °C at 10 °C/min (hold 30 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec @ 35 °C
DetectorECD @ 290 °C