Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Organochlorine Pesticides US EPA Method 8081 on Rtx-440

ColumnRtx-440, 30 m, 0.32 mm ID, 0.50 µm (cat.# 12939)
using Siltek Guard Column 5 m, 0.32 mm ID (cat.# 10027)
with SeCure “Y” Connector Kit (cat.# 20277)
SampleOrganochlorine Pesticide Mix AB #2 (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000)
Conc.: Organochlorine Pesticide Mix AB #2, 8-80µg/mL each component in hexane/toluene (cat.# 32292), Pesticide Surrogate Mix, 200µg/mL each component in acetone (cat.# 32000)
Injection
Inj. Vol.:0.5 µL splitless (hold 0.75 min)
Liner:3.5 mm Single Gooseneck (cat.# 20961)
Inj. Temp.:250 °C
Oven
Oven Temp.:110 °C (hold 0.5 min) to 230 °C at 25 °C/min to 330 °C at 15 °C/min (hold 1 min)
Carrier GasH2, constant pressure
Linear Velocity:60 cm/sec @ 300 °C
DetectorECD @ 330 °C
InstrumentShimadzu 2010 GC