Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Organochlorine Pesticides on Rtx-CLPesticides

GC_EV00904
ColumnRtx-CLPesticides, 20 m, 0.18 mm ID, 0.18 µm (cat.# 42102)
SampleOrganochlorine Pesticide Mix AB #2 (8-80µg/mL each component in hexane/toluene 1:1) (cat.# 32292)
Pesticide Surrogate Mix, EPA 8080, 8081 (200µg/mL each component in acetone) (cat.# 32000)
Injection
Inj. Vol.:0.5 µL split (split ratio 20:1)
Liner:Gooseneck Splitless (2mm) (cat.# 20796)
Inj. Temp.:250 °C
Oven
Oven Temp.:275 °C
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorECD @ 325 °C
NotesColumn Heater: Gerstel Modular Accelerated Column Heater
Gerstel module temp.: 150°C (hold 10 sec) to 190°C (hold 0.5 min.) @ 300°C/min., to 320°C @ 40°C/min. (hold 2 min.)

This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual-columns connected to a 5-meter guard column using a “Y” Press-Tight connector.