Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Organochlorine Pesticide Mix AB #2 Rtx-CLPesticides and Rtx-CLPesticides2 (0.25mm ID column set)

ColumnsRtx-CLPesticides 30 m, 0.25 mm ID, 0.25 µm (cat.# 11123)
and Rtx-CLPesticides2 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029)
with Siltek-treated universal “Y” Press-Tight connector (cat.# 20486)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Pesticide surrogate mix, EPA 8080, 8081 (cat.# 32000)
Injection
Inj. Vol.:0.5 µL splitless (hold 0.5 min)
Liner:Cyclo double taper (2 mm) (cat.# 20908)
Inj. Temp.:250 °C
Oven
Oven Temp.:125 °C to 200 °C at 45 °C/min to 230 °C at 12.5 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant linear velocity
Linear Velocity:41 cm/sec @ 125 °C
DetectorECD @ 330 °C
NotesThis chromatogram was obtained using an Agilent μ-ECD. To obtain comparable results, you will need to employ a μ-ECD in addition to dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.