Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Aroclor 1221 PCBs on Rtx-CLPesticides

GC_EV00993
ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
SampleAroclor 1221 in hexane diluted to 1,000 ppb
decachlorobiphenyl (BZ #209) in acetone diluted to 100 ppb (cat.# 32029)
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.3 min)
Liner:Cyclo Double Gooseneck (4mm) (cat.# 20895)
Inj. Temp.:250 °C
Pulse Pressure:30 psi (206.8kPa)
Oven
Oven Temp.:120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant linear velocity
Linear Velocity:60 cm/sec @ 120 °C
DetectorECD @ 330 °C
NotesThis chromatogram was obtained using an Agilent μ-ECD. To obtain comparable results, you will need to employ a μ-ECD in addition to confirmational dual-columns connected to a 5-meter guard column using a “Y” Press-Tight connector.