Mineral Oil on Rxi-5HT by EN9377-2 (PTV large volume injection)

GC_EV1180
ColumnRxi-5HT, 15 m, 0.32 mm ID, 0.10 µm (cat.# 13906)
SampleMineral oil
Diluent:Hexane
Conc.:25 ppm
Injection
Inj. Vol.:100 µL PTV splitless
Inlet Temp. Program:45 °C (hold 0.45 min) to 350 °C at 200 °C/min (hold 10 min)
Oven
Oven Temp.:35 °C (hold 4 min) to 150 °C at 60 °C/min to 250 °C at 50 °C/min to 350 °C at 30 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorFID @ 360 °C
InstrumentVarian 450
AcknowledgementAmbiente Analisi S.r.L, Italy