Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Phenanthrene and Anthracene on Rxi-17 (15m x 0.25mm x 0.25µm)

GC_EV1183
ColumnRxi-17, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13520)
Standard/Sample
Diluent:Methylene chloride
Conc.:20 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:4mm Split Precision Liner w/Wool (cat.# 21022)
Inj. Temp.:275 °C
Split Vent Flow Rate:42 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 320 °C at 15 °C/min (hold 2 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 340 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC