Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

TCDDs on Rtx-Dioxin2 Column

GC_EV1239
PeakstR (min)
1.1,2,3,4-TCDD27.92
2.1,2,3,7-TCDD28.65
3.1,2,3,8-TCDD28.65
4.1,2,3,9-TCDD28.78
5.2,3,7,8-TCDD29.40
6.1,2,6,7-TCDD29.58
7.1,2,7,8-TCDD29.58
8.1,2,8,9-TCDD30.02
ColumnRtx-Dioxin2, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10758)
SampleLaboratory prepared test mix
Injection
Inj. Vol.:1.0 µL splitless
Inj. Temp.:290 °C
Oven
Oven Temp.:150 °C (hold 1 min) to 210 °C at 30 °C/min (hold 1 min) to 250 °C at 3 °C/min (hold 12 min) to 330 °C at 70 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorMicromass Autospec Ultima
Transfer Line Temp.:290 °C
Analyzer Type: Magnetic Sector
Source Temp.:290 °C
Tune Type:PFK
Ionization Mode:EI
InstrumentAgilent/HP6890 GC
AcknowledgementCourtesy of Maxxam Analytics (Ontario, Canada).