Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Florisil SPE Cleanup of EPA 8081 Pesticides on Rtx-CLPesticides2

ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal “Y” Press-Tight connector (cat.# 20406-261)
Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Organochlorine pesticide mix AB #2 (cat.# 32292)
Diluent:Hexane
Injection
Inj. Vol.:2 µL splitless (hold 0.75 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:110 °C (hold 0.5 min) to 320 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
Detectorµ-ECD @ 330 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
NotesSPE: Condition Florisil tube (cat.# 24034) with 6 mL hexane. Elute with hexane:acetone (90:10), collecting 10 mL. Concentrate to 1 mL.