Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Tetrachlorodioxins on Rtx-Dioxin2 (60 m x 0.18 mm x 0.10 µm)

GC_EV1302
ColumnRtx-Dioxin2, 60 m, 0.18 mm ID, 0.10 µm (cat.# custom)
SampleFly ash extract
Diluent:Nonane
Conc.: This is a "real world" sample extract.
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Premium 2 mm single taper w/wool (cat.# 23316.1)
Inj. Temp.:280 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:120 °C (hold 1 min) to 200 °C at 35 °C/min to 280 °C at 4.5 °C/min to 330 °C at 20 °C/min (hold 4.8 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorHRMS
Mode:SIM
Transfer Line Temp.:300 °C
Analyzer Type:Magnetic Sector
Source Temp.:280 °C
Tune Type:PFK
Electron Energy:40 eV
Resolving Power:10,000
InstrumentWaters AutoSpec Ultima Mass Spectrometer
AcknowledgementTerry Kolic, Ontario Ministry of Environment