Florida TRPH Standard on Rxi-1HT (15 m x 0.25 mm x 0.10 µm)

GC_EV1362
PeakstR (sec)
1.C860.7
2.C10118.7
3.C12195.2
4.C14269.6
5.C16337.6
6.C18399.2
7.C20455.3
8.C22507.0
PeakstR (sec)
9.C24554.5
10.C26598.3
11.C28639.7
12.C30678.3
13.C32714.3
14.C34748.3
15.C36780.5
16.C38811.1
17.C40840.2
ColumnRxi-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleFlorida TRPH standard (cat.# 31266)
Diluent:Hexane
Conc.:500 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 250:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:437.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 20 °C/min (hold 1.9 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc).